- Design qualification (DQ)
- Installation qualification (IQ)
- Operational qualification (OQ)
- Performance qualification (PQ)
Design
elements |
Examples |
Intended use |
Analysis of drug compounds and impurities. |
User requirement specification for the HPLC
analysis |
Up to 100 samples / day. |
Automated over-night analysis. |
|
Limit of quantitation: 0.1%. |
|
Automated confirmation of peak identity and
purity with diode-array detection. |
|
Automated compound quantitation and printing of
report. |
|
Functional specifications |
Pump: Binary or higher gradient. |
Detector: UV/Vis Diode-array, 190-900 nm. |
|
Autosampler: 100 samples, 0.5 µl to 5 ml sample
volume. |
|
Column compartment: 15 to 60 Deg C, Peltier
controlled. |
|
Computer: System control, data acquisition for
signals and spectra, peak integration and quantitation, spectral evaluation
for peak purity and compound confirmation. Electronically save all
chromatograms generated by the system. |
|
Operational specifications |
Detector: Baseline noise: <5 x 10-5 AU Sampler: Precision inj. volume: <0.5% RSD, sample
carry over: <0.5% Pump: precision of retention time: <0.5% RSD |
User instructions |
Operational manual on paper. Computer-based tutorial. |
Validation/qualification |
Vendor must provide procedures and services for IQ and OQ. |
Maintenance |
Vendor must deliver maintenance procedure and
recommend schedule. Instrument must include early maintenance
feedback for timely exchange. |
- Obtain the manufacturer's recommendations for the installation site requirements.
- Check the site for the fulfillment of the manufacturer’s recommendations (utilities such as electricity, and environmental conditions such as humidity and temperature).
- Allow sufficient shelf space for the equipment, SOPs, operating manuals and software.
- Compare equipment, as received, with purchase order (including software, accessories, spare parts)
- Check documentation for completeness (operating manuals, maintenance instructions, standard operating procedures for testing, safety and validation certificates)
- Check equipment for any damage
- Install hardware (computer, equipment, fittings and tubings for fluid connections, columns in HPLC and GC, power cables, data flow and instrument control cables)
- Switch on the instruments and ensure that all modules power up and perform an electronic self-test.
- Identify and make a list with a description of all hardware, include drawings where appropriate.
- Run test sample and compare chromatogram print-out with reference chromatogram.
- List equipment manuals and SOPs
- Prepare an installation report
Parameter |
Procedure |
User Limit |
Leak testing |
Flow test by volume or weight/time |
± 5 % |
Baseline drift |
ASTM Method E19.09, 20 min |
<2 x 10-3 AU |
Baseline noise |
ASTM Method E19.09, 20 x 1 min |
<5 x 10-5 AU |
Precision of injection volume |
6 x injection of caffeine standard, RSD of peak areas |
0.3 % RSD |
Detector linearity |
inject 5 standards |
>1.5 AU, 5% |
Wavelength accuracy |
holmium oxide filter |
± 1 nm |
Temperature accuracy |
comparison with external measuring device |
± 1 °C |
Temperature precision |
monitoring temperature over 20 min |
± 0.25 °C |
Autosampler carry over |
Injection of blank solvent after large concentration |
< 0.5 % |
Mobile phase composition accuracy |
Step gradients from 4 to 7 % B, step heights
relative to 100%, with acetone tracer |
± 1 % |
- The precision of the amounts
- The precision of retention times
- Resolution between two peaks
- Peak width at half height or
- Peak tailing
- Baseline noise
- Wavelength accuracy of the UV/Visible wavelength detector, preferably using built-in holmium-oxide filters
- Design qualification document
- IQ document (includes a description of hardware and software)
- Procedures for OQ testing
- OQ test reports
- PQ test procedures and representative results
0 Comments