- Precision.
- Specificity.
- Limit of detection.
- Limit of quantitation.
- Linearity.
- Range.
- Robustness.
Checklist for Analytical Method
Validation (Chemical)
Assay/Related Substances
TEST |
ASSAY/RELATED
SUBSTANCES |
||
PARAMETER |
No. |
DOCUMENTS REQUIRED |
AVAILABILITY |
Specificity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Chromatogram/spectrum
for following solutions: |
|
|
|
Standard |
|
|
Sample |
|
||
Blank/Placebo |
|
||
Spike
solution |
|
||
System
suitability tests |
|
||
4 |
Impurities
available |
|
|
|
Peak
purity for PDA detector or spike solution for non-PDA detector |
|
|
Relative
Retention Time (RRT) |
|
||
Resolution
(R) where applicable |
|
||
5 |
Impurities
not available |
|
|
|
Stress
data (minimum data for humidity, heat and light) |
|
|
Peak
purity |
|
||
Relative
Retention Time (RRT) |
|
||
Resolution
(R) where applicable |
|
||
Linearity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
five (5) levels of standard solutions |
|
|
4 |
Data
for linear regression equation, Y-intercept, slope, r2 and linearity graph |
|
Range |
1 |
For
the assay of drug substances: 80% - 120% of working concentration
(WC) |
|
2 |
For
the determination of an impurity: from the reporting level of an impurity to 120%
of the specification |
|
|
Accuracy |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range |
|
|
4 |
Result:
reported as percent recovery by the assay of known added amount of analyte in
the sample, OR as the difference between the mean and the accepted true value
together with the confidence intervals |
|
|
Precision (Repeatability) |
1 |
Testing
Method (using sample/product as the test solution) |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range,
OR minimum six (6) replicates at 100% of the WC |
|
|
4 |
Result:
SD, RSD and confidence Interval |
|
|
Precision
(intermediate precision/ ruggedness) |
1 |
Testing
Method (using sample/product as the test solution) |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range,
OR minimum
six (6) replicates at 100% of the WC |
|
|
4 |
Cover
at least 2 parameters among variation of analyst, date and equipment |
|
|
5 |
Result:
SD, RSD and confidence Interval |
|
Detection Limit |
1 |
Testing Method: visual observation / signal-to-noise
/ standard deviation of the response and the slope |
|
2 |
If based on standard deviation of the response and
the slope method |
|
|
|
Minimum five (5) levels of standard solutions |
|
|
Peak area values for all concentrations |
|
||
Data for linear regression equation, Y-intercept,
slope, r2 and linearity graph. |
|
||
3 |
Calculation/formulation where applicable |
|
|
4 |
Related Chromatogram(s) at LOD |
|
|
5 |
Value of detection limit |
|
|
Quantitation Limit |
1 |
Testing Method: visual observation / signal-to-noise
/ standard deviation of the response and the slope |
|
2 |
if based on visual observation method, accuracy and
precision data at LOQ must be provided |
|
|
3 |
If based on calibration curve method |
|
|
|
Minimum five (5) levels of standard solutions |
|
|
Peak area values for all concentrations |
|
||
Data for linear regression equation, Y-intercept,
slope, r2 and linearity graph. |
|
||
5 |
Calculation/formulation where applicable |
|
|
6 |
Value of quantitation limit |
|
|
System Suitability
Testing |
1 |
RSD, tailing factor, theoretical plate |
|
2 |
Resolution (if two or more components) |
|
|
Robustness (not mandatory) |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Result: refer acceptance criteria for accuracy and
precision (repeatability) |
|
Checklist for Analytical Method Validation (Chemical)
Dissolution
TEST |
DISSOLUTION |
||
PARAMETER |
No. |
DOCUMENTS REQUIRED |
AVAILABILITY |
Specificity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Chromatogram/spectrum
for the following solutions: |
|
|
|
Standard |
|
|
Sample |
|
||
Blank/Placebo |
|
||
Spike
solution |
|
||
System
suitability tests |
|
||
Linearity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
five (5) levels of standard solutions |
|
|
4 |
Data
for linear regression equation, Y-intercept, slope, r2, and linearity
graph. |
|
|
Range |
1 |
Dissolution
testing: ± 20% over the specified range Example
1: if the specification is NLT 75% (Q) of the labeled amount is dissolved in
45 minutes, the validated range would be 60 – 100% of the label claim Example
2: if the specification for a controlled released product covers a region
from 20% after 1 hour, up to 90%, after 24 hours, the validated range would be
0 – 110% of the label claim |
|
Accuracy |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range |
|
|
4 |
Result: reported as percent recovery by the assay of known added amount of analyte in
the sample, OR as the difference between the mean and the accepted true value
together with the confidence intervals |
|
Precision
(Repeatability) |
1 |
Testing
Method (using sample/product as the test solution) |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range,
OR minimum six (6) replicates at 100% of the WC |
|
|
4 |
Result:
SD, RSD and confidence Interval |
|
|
Precision
(intermediate precision/ ruggedness) |
1 |
Testing
Method (using sample/product as the test solution) |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
three (3) levels of concentration in triplicates covering the specified range, OR minimum six (6) replicates at 100% of the WC |
|
|
4 |
Cover
at least 2 parameters among variations of analyst, date and equipment |
|
|
5 |
Result:
SD, RSD and confidence Interval |
|
|
System Suitability Testing |
1 |
RSD,
tailing factor, theoretical plate |
|
2 |
Resolution
(if two or more components) |
|
|
Robustness (Not Mandatory) |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Result:
refer to acceptance criteria for accuracy and precision (repeatability) |
|
PARAMETER |
ACCEPTANCE
CRITERIA |
Specificity |
Absence
of interfering peaks in the placebo, impurity demonstrates specificity |
Pass
peak purity test (particularly for related substances test) |
|
Linearity |
r2
≥ 0.995 |
y-intercept
at 100% working concentration ≤ 2% |
|
Accuracy |
Measured
recovery within 95% - 105% or mean difference ± 2% & CI |
Precision (Repeatability) |
RSD
≤ 2.0% & CI |
Precision
(intermediate precision/ ruggedness) |
RSD
≤ 2.0% & CI or mean
difference ± 2% & CI |
Detection Limit |
LOD
peak must be visible |
If
based on a standard deviation of the response and the slope method, DL = 3.3 Ó/S |
|
If
based on signal to noise, S/N= 3:1 or 2:1 |
|
Quantitation
Limit |
if
based on the visual observation method, accuracy and precision data at LOQ must be
± 20% |
If
based on a standard deviation of the response and the slope method, DL = 10 Ó/S |
|
If
based on signal to noise, S/N= 3:1 or 2:1 |
|
System
Suitability Testing |
RSD
≤ 2% |
Theoretical
plate/column efficiency, N ≥ 2000 |
|
Tailing
factor < 2 |
|
Resolution
> 2 |
|
Robustness (Not mandatory) |
Refer to acceptance criteria for accuracy and precision (repeatability) |
0 Comments