HOT POST

50/recent/ticker-posts

Header Ads Widget

SOP for Shimadzu Spectrofluorophotometer

OBJECTIVE
This SOP describes the procedure for the operation of the Shimadzu Spectrofluorophotometer RF-5301PC.

SCOPE
Spectrofluorophotometer is used for qualitative and quantitative analysis of fluorescent organic and inorganic compounds.

RESPONSIBILITIES
It is the responsibility of the QC Analyst to ensure adherence to this procedure.
It is the responsibility of the QC Manager to follow the instructions of this procedure.

PROCEDURE
Starting the Instrument
  • Turn on the computer,
  • Turn on the instrument, The on/off switch is located in the lower right corner
  • Starting the software, Double click the icon PC 150X

Software Operation
  • To open the menu, Select acquire mode drop-down menu
  • Ensure spectrum is checked
  • Select configure drop-down menu
  • Select spectrum parameter
  • Select spectrum type
  • When excitation is selected, the emission grating is in a fixed position
  • When emission is selected, the excitation grating is in a fixed position


Enter desired parameters
  • Excitation/Emission wavelength: 0, 220-900nm
  • Excitation/Emission wavelength range: 220-900nm
  • The minimum scanning range is 25
  • Sensitivity can be set to high or low
  • Recording range – limit of the graph
  • The allowable range is -100 to 1000
  • Repeat Scan - allows automatic consecutive scanning from 2-100 scans
  • Press OK.


Running an Experiment
  • Loading a Sample
  • Flip door down
  • Insert a cuvette
  • Close the door
  • Click on the start button
  • When the scan is completed the file name dialogue box appears
  • Type in a file name
  • Maximum of 8 characters
  • A comment may be typed into the dialogue box

Select presentation drop-down menu
  • Press graph
  • Press radar
  • Both axes

Zoom In
  • Click a drag a box around the area you wish to enlarge
  • To return to full screen
  • Right mouse click
  • Select limits
  • Press reset
  • Press ok


Cross Hair Coordinates
  • Right mouse click
  • Select cross hair
  • Select display
  • To remove the cross-hair coordinates press display again
  • Select presentation drop-down menu
  • Select plot to display the plot layout window
  • Under the graph column
  • Click on the first box in row
  • If it is already checked, uncheck the box then recheck the box
  • The default for plot data is a screen, this will print what appears on the screen
  • Select Ok
  • Select the box under Params

Printing a Spectrum
  • Click boxes 3 and 4 under quadrants
  • The boxes are on the right side of the menu
  • Select preview to display what will be printed

Select Print
  • Shutting down the instrument
  • Remove cuvette
  • Close PC-1501 program
  • Turn off instrument

ABBREVIATION
QC: Quality Control

ANNEXURE
Nil

REVISION HISTORY
Nil

Post a Comment

0 Comments

close