OBJECTIVE
To provide guidelines for the Operation of the Digital Bottle Cap Torque Tester.
POLICY
To establish a procedure for the operation of the Digital Bottle Cap Torque Tester.
RESPONSIBILITY
Q.A Executive.
PROCEDURE
- Ascertain that the system is properly placed on a hard & even surface. Plug the adapter into a wall electric socket.
- Place the bottle between the grips and turn the screw till the bottle is rigidly held in it.
- Switch on the system, wait for two minutes.
- Push the Zero Button.
- Press the PEAK button
- Select the unit in which torque value is required to display / measure. The torque value required for Blue Hawaii hair oil is Ncm.
- Apply the force on the cap / seal to open it manually.
- The maximum torque value will be held, if peak was selected.
- Note the reading
- To bring the system in normal condition, push the PEAK button once again.
- Repeat operation 3.2 to 3.10 for each and every bottle checked.
ANNEXURE
Nil
REVISION HISTORY
Nil