OBJECTIVE
To lay down the procedure for operation and calibration of the Top Loading Balance for better and error-free use.
SCOPE
This procedure is applicable for Balance (MAKE: Ohaus Corporation, MODEL: V31XH202)
RESPONSIBILITY
Execution: Chemist and above – QC department.
Checking: Asst. Manager and above – QC department
PROCEDURE
Operation Procedure
- Make sure that the balance is kept clean.
- Ensure that the calibration status is valid.
- Ensure that the spirit level is in the center of the circle.
- Connect the power cable to the mains and switch ‘ON’.
- Press ON/OFF key, all the display will glow.
- Press “TARE KEY”, 0.00 mark appears on display.
- Once the stability is attained, the balance is ready for weighing.
- Place the material to be weighed on the pan & note down the reading.
- Clean the balance immediately after weighing
PEOPLE ALSO READ: SOP for Cleaning Procedure for Analytical Balances
Preliminary Checks for Balance Calibration
- Spirit Level Check (Balance Levelling):
- Adjust the air bubble to the centre of the inner circle of the indicator if it is not in the centre.
- Acceptance Criteria: Air bubble must be in centre of inner circle.
- Pan Cleaning and Adjustment:
- Before placing the std. weight/butter paper/glassware, clean the pan properly
- Also ensure that the pan is adjusted correctly (if not adjusted before).
- Acceptance Criteria: Pan must be cleaned and centrally placed.
- Zero Error Check:
- Press the ON/OFF key and wait till LCD shows zero.
- Acceptance Criteria: LCD should display zero on the LCD screen.
- Handling of Weight Box:
- Ensure that the weights used are calibrated and certified on a regular basis.
- Always use forceps/tweezers to place the weight on the balance pan.
- Do not drop the weight with force on the pan of the balance, to avoid any variation in the results
- Weights used for calibration should always be cleaned with tissue paper before & after each use
- Acceptance Criteria: Weights should not have any denting or damage sign and should be cleaned before and after each use.
PEOPLE ALSO READ: SOP for Operation and Calibration Procedure For Analytical Balance
Calibration Procedure
FREQUENCY: Daily
External Calibration
ACCURACY:
- Place 1g weight on the weighing pan.
- Note down the observed weight
- Repeat the above steps using 100g and record the reading in Annexure-1
- Acceptance Criteria: Each weight should not differ by more than 0.1% of the actual weight.
ABBREVIATION
SOP : Standard Operating Procedure
No. : Number
Dept. : Department
QC : Quality Control
Asst. : Assistant
g : Gram
LCD : Liquid Crystal Display
ANNEXURE
Annexure No. - 1 -- Preliminary Checks & Calibration Record of the Balance
REVISION HISTORY
Nil