OBJECTIVE
To lay down a procedure for calibration of the hygro-thermometer.
SCOPE
This procedure applies to the production department in the Pharmaceutical Manufacturing Facilities of XYZ Limited.
RESPONSIBILITY
The IPQA staff is responsible for the implementation of the process.
The QA In-Charge is responsible for the execution and compliance of the procedure.
PEOPLE ALSO READ: SOP for Ultrasonic Bath
PROCEDURE
Measuring Range: Temperature – 10℃ to 50℃
RH – 20% to 95%.
- Collect in-house hygro-thermometer and compare it with standard hygro-thermometer.
- Standard hygro-thermometer should be validated by the external body.
- Check the date of validity of the standard hygro-thermometer.
- Keep in-house hygro-thermometer with standard hygro-thermometer in one place.
- Check the reading room in-house and standard hygro-thermometer at 9.00AM, 1.00PM, and 5.00PM and record it in the calibration log.
- The average tolerance limit is +/-5%R.H.
- Frequency of calibration: Once a Month.
- Calibration details shall be entered in Annexure-1.
ABBREVIATION(s)
Q.A : Quality Assurance
SOP : Standard Operating Procedure
No. : Number
ANNEXURE(s)
Annexure No. 1 - Calibration log for hygro-thermometer
REVISION CARD
Nil