Purpose:
To develop the standard Operation procedure for GC-HS; Agilent 6797A and its compliance with the provision of Good laboratory practice regulations.
Scope:
Describes the finest details of the steps to be followed for the simplest and most precise handling of the machine and analytical technique
Responsibilities:
All the staff members carrying out the operation of GC-HS; Agilent is responsible for strictly adhering to the procedure given.
PEOPLE ALSO READ: SOP for Calibration of HPLC
Procedure:
Cold Start of Gas Chromatography Headspace Instrument
- Before starting the Instrument, switch “On” the Air Conditioning of the room.
- Switch “On” all the gases (Nitrogen, Hydrogen & Zero air) by pushing the valve up on Gas Panel.
- Switch “On” the main UPS by long pressing start/stop button for 3 seconds. Before starting the GC Instrument, check the voltage stability (Voltage should be #233V).
- Switch “On” the GC instrument by pressing the button on the lower left corner.
- Similarly, switch “On” the HS instrument by pressing the button on the lower left corner.
- Switch “On” the CPU of the computer.
- As the system loads, from the desktop, select ‘GC’ from Administrator and GC.
Preparation of Sample and Standards for Analysis
Preparation of Standard
- From the standard GC grade analyte (99.95% GC grade), pipette 1ml and dilute the analyte in 100ml Ultra-pure water.
- Pipette out serial dilutions of 2000µl, 1000µl, 500µl, 250µl, 100µl, from stock solution of 1ml/100ml and dilute in 10ml of Ultra-pure water.
Preparation of Internal Standard
- From the standard GC grade analyte (n-propanol), pipette 300µl and dilute the analyte in 100ml distilled water.
- Take 90µl of internal standard in the vial along with Standard and Sample if required.
Sample Preparation
- In a clean, dry 20 ml HS vial take 1000µl/1ml of liquid analyte.
- If required as per analytical conditions, add 90µl of Internal Standard (n-propanol).
- Seal the mouth of vial using fresh PTFE septa and metallic crimp.
- Shake the vial slowly to mix sample and standard.
Creating New Method for Analysis
Creating New Method
- On the Desktop, Double click on the “Instrument1 online” icon; the Method and Run Control screen will appear. 2. From the Menu bar click on “Method”, from the drop-down list select “New Method”.
Creating Calibration of a Method
- Successively run samples of serial dilutions; 2000µl, 1000µl, 500µl, 250µl, 100µl in a sequence.
- Double-click on the “Instrument 1 offline” icon on the desktop; the Data analysis screen will appear.
- From the Menu bar click on “Method”, from the drop-down list select “Load Method”. From the list, select the method on which calibration is to be drawn.
- Select the graph of maximum concentration and load.
- On the “Menu bar” click on “Calibration”, from the drop-down list select “New Calibration Table”.
- “Calibrate: Instrument 1, New Calibration Table” window will appear, click on “OK”.
- All the peaks appear highlighted and table appears with RT and area.
- For every peak specify “Name” and “Amount”.
- For next dilution, from the “Menu bar” click on “Calibration”, from the dropdown list select “Add Level”.
- Specify only amount. After all concentrations have been labelled, click on “OK”.
- From the Menu bar click on “Method”, from the drop-down list select “Save Method”.
PEOPLE ALSO READ: SOP for Operation of HPLC
Operation of Gas-Chromatography Headspace System
- Double-click on the “Instrument 1 online” icon on the desktop; the Method and Run Control screen will appear.
- From the Menu bar click on “Method”, from the drop-down list select “Load Method”.
- Run the “CONDITIONING.M” method for at least 30 minutes before analyzing any sample.
Analyzing Sample on GC-HS
- From the Menu bar click on “Method”, from the drop-down list select “Load Method”. From the list, select the method to be run (recent calibration method), and click ‘OK’.
- From the Menu bar click on “Sequence”, from the drop-down list select “Sequence Table”.
- On the Sequence Table: Instrument1 window, fill the following data of
- Vial position where vial is kept
- Sample Name ID of the sample to be analyzed
- Method name Method on which sample is to be analyzed
- Inj/vial Always “1”
- Sample type Sample/ Standard
- For every next line, press down key.
- For every vial in the carousal, information should be filled up.
- Start sequence by clicking on “Run Sequence” or by pressing F6.
Viewing Report of Sample Run
- On the Desktop, Double click on the “Instrument 1 offline” icon on the desktop; Data analysis screen will appear.
- From the Menu bar click on “Method”, from the drop-down list select “Load Method”. From the list, select the method to be loaded.
- From “Data Analysis” window, on the left side of the screen, select the sequence run.
- Load the graph.
- From the Menu bar click on “Report”, click on “Print Report”.
- View report and print if required.
Shut Down of Gas Chromatography Headspace Instrument
- On the Desktop, Double click on the “Instrument1 offline” icon. From the Menu bar click on “Method”, from the drop-down list select “Load Method”. From the list, select the “COOLING.M” method.
- As soon as the Detector temperature of the GC system and the Oven temperature of the Headspace system acquires a temperature below 50°C, close the “ChemStation Online” or “Offline” windows.
- Switch “Off” the Instruments by pressing the button at the left-down corner. Push the valve down for gases (Nitrogen, Hydrogen & Zero Air) from as panel.
- Press the “Start/Stop” button on the UPS, and wait till the supply becomes zero.
- Switch “Off” the Air conditioners of the room.
PEOPLE ALSO READ: Troubleshooting Guide - HPLC
FITTING OF NEW COLUMN FOR ANALYSIS
Fitting New Column in Oven Compartment
- Open the Oven compartment, ensuring that system is cooled and not in working condition.
- Open the nut of Inlet and Detector with a spanner.
- Insert the new column within the hanger in oven. Take one end of column to be fixed in Inlet end and other at Detector bend.
- To fix the Inlet end, cut the capillary column at about a length of 1mm with the help of cutter. Insert a septa and ferrule leaving a length of 4mm-6mm from tip.
- Insert the column tip in the Inlet end, leaving a distance of 1mm from top. Seal the attachment by screwing the nut using a spanner.
- Similarly, seal the detector end by cutting the capillary column at about a length of 1mm with the help of cutter. Insert a septa and ferrule leaving a length of 4mm-6mm from tip.
- Insert the end in detector, leaving a tip of 1mm and seal using spanner.
Configuring New Column
- From the Menu bar click on “Method”, from the drop-down list select “Load Method”. From the list, select the method on which new column is to be configured.
- On the Menu bar click on “Edit Instrument Parameters”, Setup Method window will appear.
- Select 7690A (GC Sampler), different conditions of Auto sampler; Column; Oven; Inlet; Detector etc. will appear.
- Click on “Column”, and column conditions will appear.
- Click on “Show All Columns”, list of columns appears. Select your column specifying the Identification and its length etc. Click on “Install”, and specify Inlet and Outlet.
- Click on “Calibrate Column”.
- Click on “OK” and save method changes.
- Run the Conditioning method for 1 hour before running any sample on a new column.
Revision History
Nil