SOP for Calibration of Micro Controller Melting Point Apparatus

OBJECTIVE:
To provide a procedure for the calibration of micro controller melting point apparatus. 

SCOPE:
Applicable to the micro controller melting point apparatus in the QC Chemistry. 

RESPONSIBILITY:
The task shall be performed by QC Executive. 
It is the responsibility of the Manager QC to ensure that this procedure is followed accordingly. 

PROCEDURE: 
  • Set up the instrument as per the standard operating procedure.
  • Use the following reference standards to calibrate the instrument, record the observation in the Attachment-I.
          1. Vanillin 
          2. Acetanilide 
          3. Sulphanilamide 
          4. Caffeine 
  • The melting point of the reference standards should be within the standard range 
  • Standard melting point range shall be taken from COA of standard.
  • Calibration should be carried out once every six months or after maintenance of instrument 
ATTACHMENT:
Attachment 1: Calibration record of Micro Controller Melting Point Apparatus 
An Image/Link below is provided (as is) to download Annexure
Download Policy: Content on the Website is provided to you as is for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.
Password: PHARMASKY
 
ABBREVIATION:
QC- Quality Control 
COA - Certificate of Analysis