OBJECTIVE:
To provide a procedure for the calibration of micro controller melting point apparatus.
SCOPE:
Applicable to the micro controller melting point apparatus in the QC Chemistry.
RESPONSIBILITY:
The task shall be performed by QC Executive.
It is the responsibility of the Manager QC to ensure that this procedure is followed accordingly.
It is the responsibility of the Manager QC to ensure that this procedure is followed accordingly.
PROCEDURE:
- Set up the instrument as per the standard operating procedure.
- Use the following reference standards to calibrate the instrument, record the observation in the Attachment-I.
1. Vanillin
2. Acetanilide
3. Sulphanilamide
4. Caffeine
PEOPLE ALSO READ: SOP for Handling of Chemicals and Reagents
- The melting point of the reference standards should be within the standard range
- Standard melting point range shall be taken from COA of standard.
- Calibration should be carried out once every six months or after maintenance of instrument
ATTACHMENT:
Attachment 1: Calibration record of Micro Controller Melting Point Apparatus
Download Policy: Content on the Website is provided to you as is for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.
Password: PHARMASKY
ABBREVIATION:
QC- Quality Control
COA - Certificate of Analysis